Onicon F-3100 Series Manuel d'utilisateur Page 11

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 41
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 10
11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax (727) 442-5699 • [email protected]
F-3100 Flow Meter Manual 06/14 - 0670-7 Page 11
SECTION 2.0: UNPACKING
The F-3100 is generally shipped in one package unless optional hardware or equipment is ordered.
Notify the freight carrier (all products are shipped insured) and ONICON’s Customer Service
Department if any items are damaged in transit.
2.1 CHECKING THAT YOU HAVE RECEIVED EVERYTHING
Standard Documentation
Enclosed with each F-3100 is a comprehensive documentation package that includes the
following items:
Certicate of Calibration
The F-3100 Installation and Operation Guide
Please notify ONICON if any of these items are missing.
1500 North Belcher Road, Clearwater, FL 33765 • Tel (727) 447-6140 • Fax (727) 442-5699
www.onicon.com • [email protected]
F-3100 SERIES
In-line Electromagnetic Flow Meter
Installation & Basic Operation Guide
03-11
0670-1
METER DATA CALIBRATION of PRIMARY FLOW ELEMENT
Meter Tag:
Model:
Serial Number: 220177
Component S/N's:
04N1420 38M6901
Meter Size: 3"
Max. Operating Pressure:
225 psi
Max. Operating Temperature:
212 °F
Connections: ANSI 150# Class Flanges
Peripheral device serial number 220177
Analog Flow Range: 4-20 mA = 0 to 230 Gal/Min
Full Scale Frequency Output: 200.00 Hz
Scaled Pulse Output: 1 pulse = 100 Gallons
GL Date:
4/12/2011
5/25/2011
This meter was manufactured for ONICON Incorporated.
The original manufacturer certifies that this flow meter was
calibrated against standards that are traceable to
SIT, Italy.
1500 N Belcher Road, Clearwater, Florida 33765 Tel (727)447-6140 Fax (727)442-5699
Programmed By:
FACTORY PROGRAMMED OUTPUT SIGNALS
(Performed at ONICON Factory; can be reprogrammed in the field)
Primary Calibration Date:
F-3103-111
FLOW METER
CERTIFICATE OF CALIBRATION
CALIBRATION & CONFIGURATION DATA for F-3000 SERIES MAGNETIC FLOW METERS
CT-WH -1.0951
WaterMedium:
Ka factor:
Vue de la page 10